Aisys CS²

Sample gas return port

8-14
Scavenging from an auxiliary manual breathing
circuit
Scavenge the exhaust if an auxiliary manual circuit is used with N2O
or volatile anesthetics.
An auxiliary inlet is available on the AGSS. It provides a female
connection for use with a 30 mm - 30 mm male connector (or 30 mm
- 19 mm male connector) under the breathing system. Do not use
these connectors as an outlet for exhaust flow.
A separate hose is needed to connect the exhaust from the auxiliary
manual breathing circuit to the auxiliary inlet on the AGSS.
Scavenging a gas monitor sample flow
Sample gas from a gas monitor can be scavenged by connecting the
exhaust from the monitor to either the sample gas return port located
on the front of the machine or the tubing connector provided on the
AGSS.
To scavenge from a gas monitor using the sample gas return
port, connect the exhaust from the monitor to the sample gas
return port located on the front of the machine.
To scavenge from a gas monitor using AGSS, connect tubing
from the monitor to the tubing connector on the bottom of the
AGSS underneath the breathing system.
Connect the sample gas exhaust tube to the gas return port. Exhaust
gas will be directed to the scavenging system.