GE ApexPro Service Manual page 79

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Revision C
Installation: IP Address Configuration
6.
Enter the Unity Network RX network IP address, subnet mask and default
gateway settings in the Receiver (RX) Address IP Address, Subnet and Gateway
fields.
7.
Set the Administration Port Address configuration by completing one of the
following options:
To set a fixed IP address for the administration port, enter the administrator
Port IP address, subnet mask and default gateway settings in the Admin
Port Address IP Address, Subnet and Gateway fields.
To have the administration port IP address dynamically allocated when the
ApexPro Telemetry Server starts up, select the DHCP checkbox.
8.
Click Submit.
9.
For the new IP Address settings to take effect, restart the ApexPro Telemetry
Server according to
"Stop/Start/Restart ApexPro Telemetry Server"
29.
ApexPro Telemetry Server
2001989-178
on page 2-