GE ApexPro Service Manual page 189

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Revision C
Parts Lists, Drawings and Replacement: Disassembly/Assembly
8.
Disconnect the temperature sensor wiring harness plug from the receptacle at
the back right edge of the SNMP.
9.
Using a phillips screwdriver, remove the screw securing the temperature sensor
at the front left corner of the chassis floor, behind the left edge of the chassis
fan.
127(
Note the position of the temperature sensor DIP switches for re-assembly.
10. Remove the temperature sensors and the associated temperature sensor wire
harnesses, cutting/removing any ties as required.
11. Reverse the steps in this procedure to install the temperature sensors.
12. If applicable, perform the
ApexPro Telemetry Server
2001989-178
Number
1
2
3
4
"Electrical Safety Tests"
Position
OFF
OFF
ON
ON
on page 4-10.