Rack Mounting - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Rack Mounting

Revision C
Installation: Mounting
:$51,1*
PERSONAL INJURY/EQUIPMENT DAMAGE — If the
ApexPro Telemetry Server will be installed in a rack, observe the
following guidelines associated with rack stability:
Ensure both the top and bottom of the rack are securely
fastened to a suitable and substantial building support.
Ensure the rack weight is supported by the leveling jacks and
that the rack leveling jacks are in contact with the floor.
If it is a single rack installation, ensure stabilizers are
attached.
If it is a multiple rack installation, ensure the racks are tightly
secured together.
Do not extend more than one component outside the rack at a
time.
Do not exceed the rack manufacturer's maximum load
carrying capacity specification.
Personal injury or equipment failure could result from not
following these instructions.
&$87,21
PERSONAL INJURY/EQUIPMENT DAMAGE — When using
the center position, it is important to secure the side mounting
brackets as close to the ApexPro Telemetry Server center-of-
balance as possible. Installing the brackets off-balance-center
could cause the ApexPro Telemetry Server to interfere with other
servers in the rack or failure of the installation brackets.
There are 3 different rack mounting options:
Side mounting
Shelf mounting
Rail mounting
Rack Side Mounting
&$87,21
LIFTING HAZARD — Rack mounting the ApexPro Telemetry
Server in the center position is a two-person job. Do not attempt to
install the ApexPro Telemetry Server in a rack without help. One
person needs to hold/align the ApexPro Telemetry Server in the
rack while another person secures it in place.
ApexPro Telemetry Server
2001989-178
3-13