Alarms 671 - 699 - Siemens Maxum II Manual

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

#
Text
511
!
Program Failed event # %3
%4
512
? Program execution cancel‐
led: event # %3
513
!
Program Failed: Run re‐
quested on running event #
%3
514
!
Program Invalid frequency;
disabling event # %3
515
? Program Overrun for event
# %3
516
!
Formula Failure: %3
561
+ EZChrom download
562
+ EZChrom upload for app
%1 method %3
4.6.5

Alarms 671 - 699

GCP Alarm Descriptions 671 - 699
The following tables list the alarm number (#), type (+ information, ? warning, ! error) alarm
text, description, and actions.
#
Text
671
!
Database: Failure: %3
672
!
Database: Remote Service
lost on %3
673
!
Database:value > limit: %3 Limit exceeded.
674
!
Database value < limit: %3 Limit exceeded.
675
!
Database: No Stream at cy‐
cle start on applicaton %1
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
Error running MaxBasic program.
Cancellation requested from HMI or CIM
Display.
Occurs when overrun_option is set to 2
and event is run while event is still run‐
ning from a previous request.
Invalid program frequency.
Occurs when overrun_option is set to 1
and warns that a program is running
when a previous run of the same pro‐
gram has not finished.
The condition given in the alarm text pre‐
vented the formula from returning a re‐
sult.
Informational message.
Informational message.
Description
1. Cannot find method;
2. Cannot find MaxBasic program;
3. Invalid stream for program;
4. Bad status on external result.
A connection for remote I/O or result
transmission has been closed. This
alarm is normal when the remote unit be‐
comes unavailable.
Cannot locate stream to start.
General Maintenance and Troubleshooting
Action
Check message and program. If this
program was written by Siemens, con‐
tact Customer Support.
Informational. No action necessary.
Check event timing or change overrun
option.
Check program_schedule setup.
Check cycle event timing or frequency
of program or ignore.
Investigate the condition given in the
alarm text.
No action necessary.
No action necessary.
Action
Check methods and sequences.
Check program table.
Check program streamcontext.
Verify extresult table entries.
Check the status of the remote unit.
See Alarm Text.
See the Limits and Alarm Handlers ta‐
bles under the GCP Application View.
See Alarm Text.
See the Limits and Alarm Handlers ta‐
bles under the GCP Application View.
Check sequence to make sure that the
entries are enabled.
41