General Maintenance and Troubleshooting

4.5 Accessing the Bootloader

4.5
Accessing the Bootloader
Accessing the Bootloader to Set the Network Address
This procedure is needed when a CAC3 is replaced, when the memory backup battery is
removed, or when the CAC3 is removed from a CIM or SIB3. In this procedure, the term
"SYSCON" can mean either a SYSCON assembly in a Maxum Airless/Airbath model or a CIM
in a MAxum Modular Oven model analyzer. To prepare, disconnect the analyzer from the
network by unlugging the Ethernet cable.
From the color touchscreen in the door of the analyzer, use this set of steps to set the IP
addresses. This allows the GCP software to communicate with the device so that a database
may be loaded.
1. Push the reset button through the opening on the SYSCON cage. This provides access to
the bootloader.
2. Press the Home key when the message "To enter the bootloader menu, press the Home
key now..." appears.
3. Press the 5 key to select the "Configuration" menu.
4. Press the 4 key to select "Choose Device Mode".
5. Press the 2 key to select "Standalone SysCon".
6. The SYSCON reboots into standalone mode.
7. Press the Home key to re-enter the boot loader
8. Press the 5 key to select the "Configuration" menu again.
9. Press the 2 key to select "Primary Ethernet IP Configuration".
10.The question "Do you want to change this configuration?" appears: Press the 9 key to select
"Yes".
11.Press the 0 key to disable DHCP.
12.Press the 9 key to change the IP address.
13.Enter the desired IP address. Press the "Home" key to return to the menu.
14.Press the 9 key to change the subnet mask.
15.Enter the correct subnet mask. Press the "Home" key to return to the menu.
16.Press the 9 key to change the Default Gateway address.
17.Enter the Default Gateway address. Press the "Home" key to return to the menu.
18.Press the 0 key to select "No" to the question "Change the DNS?"
19.Press any key to continue.
20.Press the Back key to return to the bootloader menu.
21.Press the 9 key to reset the device.
The external Ethernet cable may now be reconnected to the network, and the analyzer should
be visible in the GCP Network list.
30
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001