Max Input
Equipment
Pulse/function generator (HP 8116A)
Oscilloscope (HP 54121T)
Attenuator 20 dB, 2 W (4 x)
Cable Assembly BNC (5 x)
BNC T-connector
Test Setup
Procedure
1. Connect the equipment as shown in the setup figure
2. Set up the HP 8112A as follows:
Trigger Mode
Trigger Slope
Control Mode
Transition
P E R
DEL
WID
HIL
3. Set up the external pulse generator (HP 8116A) as follows:
Frequency
Width
Amplitude
Offset
4. Using the oscilloscope, adjust the HP 8112A trigger level to allow
triggering from the external pulse generator.
5. Verify that each external trigger pulse generates one complete
output cycle as shown shown here.
Figure 8-12. Trigger, Gate,E.WID & E.BUR Verification Test
TRIG
Fixed
2.0 ps
65 ns
1.0 /Ls
1.00 v
0.00 v
50 kHz
5.0 ps
2.0 v
0.00 v
Testing Performance 8-19