Trigger,Gate,E.Width
and E.Burst
Verification Test
Characteristics
8-18 Testing Performance
Figure 8-11. Pulse Performance Characteristics
Trigger
Minimum
500 mV (p-p)
amplitude
Minimum
10 ns
pulse width
Each active input generates one output pulse.
Gate
External signal enables Period Generator.
n
First output pulse synchronous with external trigger.
n
External Width
Restoration of external signal with selectable transition times and
output levels.
External Burst
Each active input transition generates a preprogrammed number of
pulses (1 to 1999). Minimum time between bursts is 100 ns.
50%
Amplitude