Output Levels Performance Test; Specifications (Values In Parenthesis Into Open Circuit); Equipment - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Output Levels
Performance Test
Specifications (values
in parenthesis into open
circuit)

Equipment

lest Setup
Procedure
8-12 Testing Performance
High Level Range
-7.90 V to 8.00 V (-15.8 V to 16.0 V)
Low Level Range
-8.00 V to 7.90 V (-16.0 V to 15.8 V)
Level Accuracy
Repeatability
Factor 4 better than accuracy
Digital Voltmeter (HP 3458A)
n
n
H P 8112A
Figure 8-8. Output Levels Performance Test
1. Connect the equipment as shown in the setup figure. Use a 50 R
feedthrough termination 0.1%.
2. Set up the HP 8112A as follows:
Input Mode
Control Mode
Transition
P E R
DEL
D T Y
HIL
3. Set up the DVM as follows:
Function
DCV
50 Ohm Feedthrough connector
and BNC to dual banana plug adapter
Voltmeter (HP 3458A)
Connected to EXT TRIGGER on rear of DVM
NORM
Off
Fixed
100 ms
6.5 ms
50%
0.00 v