Application Of X-Y Operation - Kenwood CS-1022 Instruction Manual

Hide thumbs Also See for CS-1022:
Table of Contents

Advertisement

APPLICATION OF X-Y OPERATION
CS-1022, 1021, 1012
CS-1020, 1010
* Phase Shift Measurement
A method of phase measurement requires calculations bas-
ed on the Lissajous patterns obtained using X-Y operations.
Distortion
due
to
non-linear
amplification
also
can
be
displayed.
A sine wave
input is applied to the audio circuit being
tested. The same sine wave input is applied to the vertical
input of the oscilloscope, and the output of the tested cir-
cuit is applied to the horizontal input of the oscilloscope.
The amount of phase difference between the two signals
can be calculated from the resulting waveform.
To
make
phase
measurements,
use
the following
pro-
cedure.
1. Using an audio signal generator with a pure sinusoidal
signal, apply a sine wave test signal at the desired test
frequency to the audio network being tested.
. Set the signal generator output for the normal operating
level of the circuit being tested. If desired, the circuit's
output may be observed on the oscilloscope. If the test
circuit
is overdriven,
the
sine
wave
display
on
the
oscilloscope
is clipped
and the signal
level must
be
reduced.
. Connect the channel 2 probe to the output of the test
circuit.
. Select X-Y operation by placing the TRIG MODE switch
in the X-Y position.
. Connect the channel 1 probe to the input of the test cir-
cuit.
(The input and output test connections to the vertical
and horizontal oscilloscope inputs may be reserved.)
. Adjust the channel 1 and 2 gain controls for a suitable
viewing size.
. Some typical results are shown in Fig. 26.
If the two signals are in phase, the oscilloscope trace is a
straight diagonal line. If the vertical and horizontal gain
are properly adjusted, this line is at a 45° angle.
A 90°
phase sift produces a circular oscilloscope pattern.
Phase shift of less (or more) than 90° produces an ellip-
tical oscilloscope
pattern.
The amount
of phase shift
can be calculated from the oscilloscope trace as shown
in Fig. 25.
Fig. 25 Phase shift calculation
* Phase Shift Measurement
Phase measurements
may be made with an oscilloscope.
Typical
applications
are
circuits
designed
to produce
a
specific
phase
shift,
and
measurement
of phase
shift
distortion
in audio
amplifiers
or other
audio
networks.
Distortion due to non-linear amplification is also displayed
in the oscilloscope waveform.
A sine wave
input is applied to the audio
circuit being
tested. The same sine wave input is applied to the vertical
input of the oscilloscope, and the output of the tested cir-
cuit is applied to the horizontal input of the oscilloscope.
The amount of phase difference between the two signals
can be calculated from the resulting waveform.
To
make
phase
measurements,
use
the following
pro-
cedure.
1. Using an audio signal generator with a pure sinusoidal
signal, apply a sine wave test signal at the desired test
frequency to the audio network being tested.
. Set the signal generator output for the normal operating
level of the circuit being tested. If desired, the circuit's
output may be observed on the oscilloscope. If the test
circuit
is overdriven,
the
sine
wave
display
on
the
oscilloscope
is clipped
and the signal level must
be
reduced.
. Connect an external horizontal input cable from the out-
put of the test circuit to the EXT TRIG INPUT jack of the
oscilloscope.
. Set the TRIG
MODE
switch to X-Y position for X-Y
operation.
. Connect the VERT. INPUT probe to the input of the test
circuit.
(The input and output test connections
to the
vertical
and
horizontal
oscilloscope
inputs
may
be
reversed. Use the higher vertical gain of the oscilloscope
for the lower level signal.)
. Adjust the vertical and
horizontal
gain controls
for a
suitable viewing size.
. Some typical results are shown in Fig. 26.
If the two signals are in phase, the oscilloscope trace is a
straight diagonal line. If the vertical and horizontal gain
are properly adjusted, this line is at a 45° angle.
A 90°
phase sift produces a circular oscilloscope pattern.
Phase shift of less (or more) than 90° produces an ellip-
tical oscilloscope
pattern.
The amount
of phase shift
can be calculated from the oscilloscope trace as shown
Fig. 25 Phase shift calculation
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cs-1021Cs-1012Cs-1020Cs-1010

Table of Contents