5-6.
EQUIPMENT REQUIRED
5-7. Table 1-5, of the Operating and Programming Manual, lists the equipment required for the adjustment pro-
cedures. If the test equipment recommended is not available, other equipment may be used if its performance
meets the critical specifications listed in Table 1-5. The specified equipment required for each adjustment is
referenced in each procedure.
5-8. FACTORY-SELECTED COMPONENTS
5-9. Factory-selected components are identified on the schematics and parts lists by an asterisk (*) which fol-
lows the reference designator. The nominal value or range of the component is shown. Manual change sheets
will provide updated information pertaining to selected components.
5-10. RELATED ADJUSTMENTS
5-11. Adjustments that interact are noted in the adjustment procedures. Table 5-1 lists the adjustmerrt proce-
dures and the recommended order of performance. Table
5-2,
Post-Repair Adjustments, lists the recommended
adjustments to perform following a repair to different sections of the Counter.
Table 5-2. Post-Repair Adjustments
Input Amplifier
DAC
Measurement
Executive
HP-IB
Time Base
Power Supply
Front Panel
MATE (Option 700)
Channel C (Option 030)
ADJUSTMENTS (paragraph)
FUNCTIONAL BLOCK REPAIRED
Figure 8-20
Figure 8-21
Figure 8-22
Figure 8-22
Figure 8-23
Figure 8-24
(standard oscillator)
Figures 8-24 and 8-27
(Option 010)
Figure 8-24
Figure 8-28
Figure 8-25
Figure 8-26
SCHEMATIC SHEET
5-16 through 5-20
5-1 3, 5-1 4
None
None
None
5-15 through 5-20
5-15 through 5-20
5-1 2 through 5-20
None
None
5-2 1
5-12. POWER SUPPLY VOLTAGE ADJUSTMENT
Reference: Figure
8-24
Description: The +3V dc power supply is adjusted for +3.00V dc
?
.02V dc at Test Point
"
+3" using a digital
multimeter.
HP 5334B
-
Service Manual
5-3