3. Connect the function generator signal to the HP 5334B Input A using a 50n feedthrough connector as
shown in
Figure
4-2.
VERIFY: The Counter displays 10 Hz
+
0.03 Hz.
4. Record the measurement on the Performance Test Record Card, line 1.
5.
Set the function generator as follows:
Frequency
...............................................................................................................
20 MHz
VERIFY: The Counter displays 20 MHz
+
0.3 Hz.
6. Record the measurement on the Performance Test Record Card, line 2.
7.
Set the function generator as follows:
Frequency
...............................................................................................................
30 Hz
8. Set the HP 5334B as follows:
...........................................................................................................................
AC
ON
VERIFY.
The Counter displays 30 Hz
+
0.03 Hz.
9.
Record the measurement on the Performance Test Record Card, line 3.
10. Set the function generator as follows:
...............................................................................................................
Frequency
20 MHz
VERIFY: The Counter displays 20 MHz
+
0.3 Hz
11.
Record the measurement on the Performance Test Record Card, line 4.
12. Set the function generator as follows:
Frequency
...............................................................................................................
1
MHz
NOTE
Remove the 50R feedthrough connector.
13. Set the HP 5334B as follows:
..........................................................................................................................
50R
ON
VERIFY:
The Counter displays
1
MHz
*
0.04 Ilz.
14. Record the measurement on the Performance Test Record Card, line 5.
HP 5334B
-
Scrvicc Manual
4-15