g. To check the HP-IB BLOCK using Signature Analysis, connect Pod Leads from the signature
analyzer to the A1
M a i n
Board as follows:
Signature Analyzer
~ 1 7
Test Points
START LEAD
ST (TP14)
STOP LEAD
SP (TP15)
CLOCK LEAD
SA CLK (TP16)
GROUND LEAD
Ground (TP8)
1 .
Apply power to the signature analyzer and observe that its GATE light is flashing.
2.
Set the signature analyzer pushbutton switches as follows:
START BU'ITON
Rising Edge
STOP BUTTON
Falling Edge
CLOCK BUTTON
Falling Edge
3.
Take signatures. Signatures should match those shown in Figure 8-14.
HP 5334B
-
Service Manual
8-78