3. Remove the equipotential bonding cable
Note: The cable for equipotential bonding
module
Figure 7-3
4. Loosen the eight Torx screws
5. Check the flat seal
Replace the flat seal
6. Swivel the analyzer module
7. Secure the analyzer module
– Use eight Torx screws on the base of the housing
– Use two Torx screws on the rear panel with a torque of 2.5 Nm.
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB
⑥
is installed.
Unlatching plug-in connectors at the cable equipotential bonding
NOTICE
Damage to the device through pulling the latched cable
Cables for equipotential bonding are equipped at their two ends with latching plug-in
connectors. Pulling on a latched cable can cause the plug-in connector to be torn off or its
contacts to be damaged.
• Before pulling out the plug connector, loosen the latch by pressing with your index finger.
③
for damage.
③
Installing / removing and connecting analyzer and option modules
7.4 Installing analyzer modules, standard version
④
from the blank plate of the respective slot.
④
is no longer required after the analyzer
⑤
and remove the blank plate at the corresponding slot.
if it is damaged or porous.
⑥
forward and insert it into the wall housing
⑥
in the slot as follows:
①
⑤
with a torque of 0.8 Nm.
.
69