9 Remove And Install Hand Switch Cable - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
5 9
Remove And Install Hand Switch Cable
ILLUSTRATION 5-6
HAND SWITCH CABLE
LINE CABLE UP
PARALLEL TO EDGE
OF COVER BEFORE
TIGHTENING DOWN
CLAMP
To remove the hand switch cable
1. Disconnect the cable from the connector on the body of the AMX-4+.
2. Remove the 6-32 screw securing the cable clamp, then remove the clamp and
star washer.
To install the hand switch cable, perform the above steps in reverse order. Be sure to
align the cable to the cover as indicated in Illustration 5-6.
STAR
WASHER
P CLAMP
6–32 SCREW
COIL CORD
5-8
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
RIGHT SIDE COVER
HANDSWITCH LABEL
AMX-4+ SERVICE