GE AMX-4+ Technical Manual page 54

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
Note:
Note:
4 3 2
Safety Cable
CAUTION
Note:
6. Remove screws holding cable retainer clip from top of counterweight using a
magnetic nut driver or screwdriver.
The positioning is awkward and the screw can be easily dropped.
7. Push main cable ball end out from counterweight keyhole slot.
8. Pull main cable stud end through top of column.
9. Remove old cable.
10. Push new main cable stud end through top of column over top of upper pulley.
11. Push ball end into counterweight slot.
12. Install cable retainer clip.
13. Thread new main cable through back of carriage.
14. Raise vertical arm to comfortable working height.
15. Apply Loctite to threads of cable stud. Install jam nut on cable stud.
16. Install cable stud through carriage horizontal bar and install holding nut.
17. Set cable to tension in accordance with Section 4 3 4. Tighten cable jam nut.
Do not allow cables to twist while setting tension.
1. Loosen locking nut on top of stud end of safety cable. Safety cable is not in front
of follower cable.
After next step the counterweight safety mechanism is temporarily inactive
until the new safety cable is installed.
2. Remove holding nut on bottom of stud end of safety cable.
3. Remove cable stud from carriage cable mounting strut.
4. Move horizontal arm to bottom of column.
D
This raises the counterweight to the top of the column.
5. Remove screws holding cable retainer clip from top of counterweight using a
magnetic nut driver or screwdriver.
The positioning is awkward and the screw can be easily dropped.
6. Push safety cable ball end out from counterweight groove.
7. Pull safety cable stud end through top of column.
4-6
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2169360 series2236420 series2275938 series

Table of Contents