GE AMX-4+ Technical Manual page 26

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
ILLUSTRATION 1-6
FIELD LAMP ADJUSTMENT
CAUTION
Note;
15. Examine the last exposure and determine which direction to move the field lamp
to best position it around the x-ray field.
16. Turn on the field lamp.
High Temperature and Brightness!
Components are hot in the high energy field lamp adjustment area. Do not
look directly at the field lamp. Its brightness is blinding.
17. The field lamp adjustment area will be hot so use a metal tool to move the bracket
when repositioning the light field area on the film cassette.
18. When the light field area has been repositioned equidistant around the x-ray
field, tighten the adjustment screws. Alternately tightening the two screws will
help prevent the bracket from slipping out of place.
19. Repeat steps 1 through 16 until the percentage of SID" requirements are ob
tained.
If the percentage of SID requirement cannot be obtained, the light field is
either too large or in rare instances too small. In this case the collimator will
have to be returned to the factory.
20. Replace the back cover and fasten it in place with four metric Phillips head
screws. See Illustration 1-4.
21. After any adjustment to the field light area, the collimator crosshair window posi
tion should be checked and corrected if necessary. Refer to Section 1-5.
1-8
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
ADJUSTMENT SCREWS
AMX-4+ SERVICE