GE B20 User Manual page 40

Patient monitor
Hide thumbs Also See for B20:
Table of Contents

Advertisement

Setting time zone
Time Zone menu is enable only when monitor is not connected to
network and discharge paitent/haven't admitted patient.
Monitor Setup
1. Press
key.
2. Select Install/Service and enter the password.
3. Select Time Zone
Daylight Saving: to turn on, turn off or set auto adjust
daylight savings time (DST).
DST Offset Hour: to set the hours offset for daylight savings.
DST Offset Minute: to set the minutes offset for daylight
savings.
DST Begins: to set daylight savings begin time.
DST Ends: to set daylight savings end time.
Time Sync: to set time sync options, Unity is time sync to
Unity Network, NTP is time sync to NTP Server.
NTP Config: to set NTP Server configuration.
- GMT Offset Hours: to set offset hours from GMT
(Greenwich Mean Time).
- GMT Offset Minutes: to set offset minutes from GMT.
- Primary Server IP: to set primary NTP Server IP address.
- Secondary Serv IP: to set secondary NTP Server IP address.
36
NOTES:
DST Offset Hour and DST Offset Minute menu are enable
only when the Daylight Saving is set to ON or AUTO.
DST Begins and DST Ends menu are enable only when the
Daylight Saving is set to AUTO.
Time Sync menu is shown only when monitor configuration
is Unity+HL7.
NTP Config menu is shown only when monitor configuration
is Unity+HL7 or HL7. NTP Config menu is enable only when
Time Sync is set to NTP.