Check SpO2 probe
−
SpO
: Check the sensor and connections.
2
•
Check sample gas out
−
Gases: Remove blockage from the sample gas outlet.
•
Condition Battery A, Condition Battery B
−
Condition the battery according to the instructions of the
external charger.
•
Creating OCRG snapshot
−
In neonatal mode, the value of HR, SpO
of limits.
•
Default settings returned.
−
Check your power and reset the settings.
•
DEMO MODE
−
DEMO mode is active.
•
ECG module error
−
Contact authorized service personnel.
•
End of 20 min trend data
−
Check the graphical trends and move the cursor.
•
Entropy cable off
−
Connect the Entropy cable to the Entropy module.
•
Entropy sensor off
−
Check that the sensor is properly attached to the patient.
or Apnea time is out
2
125
•
Entropy sensor check failed
−
Check sensor placement and attachment.
−
Press each electrode in the sensor.
−
Replace the sensor.
•
Frame temperature high
−
Make sure there is sufficient ventilation.
−
If the problem persists, contact authorized service personnel.
−
Turn off the monitor, wait for it cool down.
•
Freq PVCs
−
Check the patient status.
•
Incompatible SpO2 Probe
−
Replace the sensor.
−
If the problem persists, contact authorized service personnel.
•
InvBP's not zeroed
−
Zero the channel indicated or zero both channels.
•
Leads off
−
ECG: Reconnect the disconnected trunk cable, electrode or
leadwire. Change the trunk cable, lead set and module.
•
License invalid
−
Contact authorized service personnel to set up the monitor's
right license again.
•
Loading from network
−
Loading mode from network.
•
Loading failed
−
Load mode from network failed.