GE SpO
Sensor
2
TS-E-D, TS-E2-GE, TS-E4-GE
TS-SE-3
TS-F-D, TS-F2-GE, TS-F4-GE, TS-SA-D, TS-SA4-GE
TS-W-D
TS-AP-10, TS-AP-25
TS-AF-10, TS-AF-25
* Because SpO
measurements are statistically distributed, only about 2/3 of the measurements can be expected to fall within
2
by a CO-oximeter
Test methods used to establish SpO
hypoxia studies with healthy non-smoking adult volunteers over the specified SpO
samples measured by CO-oximetry. Subjects comprised both adult men and women and spanned a range of skin pigmentation.
** GE TruSignal technology have been validated for low perfusion SpO
simulator with 0.3% signal amplitude.
Table 1: Accuracy for sensors (Arms) *
Non motion
(70-100%)
±3 digits
±2 digits
±2 digits
±2 digits
±2 digits
±2 digits
accuracy: GE TruSignal SpO
measurement have been validated for no motion and motion accuracy in a controlled
2
2
2
140
Motion
(70-100%)
unspecified
unspecified
unspecified
unspecified
±3 digits
±3 digits
range. SpO
readings were compared to SaO
2
2
accuracy over the specified range in a bench top testing against BioTek Index 2 patient
Low perfusion
Neonatal
(70-100%)**
(70-100%)
unspecified
unspecified
unspecified
±3 digits
±3 digits
unspecified
unspecified
unspecified
unspecified
unspecified
unspecified
±3 digits
±
1 Arms of the value measured
values of drawn blood
2