Additionally, on the CARESCAPE network, the monitor supports
remote parameter and waveform configurations from the central
station.
NOTES:
The Entropy parameter is not supported by CIC pro and
CARESCAPE Central station.
On the iCentral, the care area settings can be setup to "Non OR"
or "OR" in the DRI service menu. Please consult the authorized
service personnel.
Non OR:
Displays up to 72 hours of trend data.
NOTE: Anesthesia Gas trends are not supported
Does not display Entropy alarm limits
OR:
Displays up to 24 hours of trend data
24
Connection to Network
Ensure that the appropriate network infrastructure is in place prior
to the installation of the patient monitor.
NOTE: The network installation and configuration are performed by
authorized service personnel only.
MC and S/5 Network
Use the CAT-5 network cable to connect the monitor to the network.
1. Make sure that the power is switched off.
2. Connect one RJ-45 connector to the network port at the back of
the monitor.
3. Connect the other RJ-45 connector to the corresponding port on
the wall.
4. Turn on the power. Confirm that the network symbol and
'Network: xxx' message are displayed.