Read and record the current leakage indicated on the tester.
(10) Change leakage tester switches to:
(11) Read and record the current leakage indicated on the tester.
(12) Change leakage tester switches to:
(13) Read and record the current leakage indicated on the tester.
(14) Set the power switch of the device under test to OFF.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.

2.8.1 Acceptance criteria NC

100 microamperes (0.1 volts on the tester), and the device under test is powered from
100-240 V/50-60 Hz
2.8.2 Acceptance criteria SFC ñ ground (earth), line or neutral open
(USA only) 300 µA, and the device under test is powered from 100-120 V/50-60 Hz
(USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
500 µA, and the device under test is powered from a non-centertapped, 200-240 V/50-60
Hz, single-phase circuit
NOTE: If the reading is greater than the specification below, and the device under test is
powered from 100-240 V/50-60 Hz, the device under test fails. Contact GE Technical Support.
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Electrical Safety Tests
3-9
Document no. 2044677-001