B30 Patient Monitor
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The B30 is intended for use in the electromagnetic environment specified below. The customer or the user of
the B30 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transients/bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
B-2
.
Document no
2044677-001
Guidance and manufacturer's declaration – electromagnetic
immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If user of the equipment
requires continued operation during
power mains interruptions, it is
recommended that the equipment be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic field
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.