Recommended Tools; Recommended Parts; Inspection - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
The instructions include a check form
the procedures.
The symbol
performing the procedure.

4.2.1 Recommended tools

Tool
Screwdriver
Ambient pressure manometer
Flowmeter
Flow cassette 50/1.1
Calibration gas and the regulator 755580* (gas)
E-PSMW / E-PSMPW
Patient simulator
NOTE: *)Ensure that the calibrating gas and regulator are functioning properly before
calibration. Perform annual maintenance on the regulator as required. For more information
see
"Calibration gas regulator flow check"

4.2.2 Recommended parts

Parts
Recorder paper
Detach the module box cover by removing the five screws from the locking side of the
module. Be careful with the flex cable, when opening the cover. Remove the grey plastic
shield over the miniC unit.

4.2.3 Inspection

General
1.
Check internal parts
N-F(C)REC:
Open the paper compartment hatch and take out the paper roll, if installed.
8-12
.
Document no
2044677-001
in the instructions means that the check form should be signed after
all screws are tightened properly
all cables are connected properly
tubes are not pinched and there are no sharp bends on them
all tubes are connected properly
EMC covers are attached properly
there are no loose objects inside the module
("APPENDIX
E:") which should be filled in when performing
Order No.
Notes
Pozi-drive, Torx T10
Amb. Press. can be checked
from local meteorological
station
TSI model 4140 recommended
873812
Contains 5 CO
755534
in the
"Gas calibration"
Order No.
Notes
74205
and air
2
chapter.