GE ApexPro Operator's Manual page 210

Telemetry system
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Technical specifications
Physical specifications
FCC compliance information
Certifications
E-4
ECG gain accuracy
Common mode rejection
Defibrillator protection
Defibrillator recovery time
Pacemaker detection
Patient leakage current
Serial communications
Height
Width
Depth
Weight
This device complies with Part 95 of the FCC Rules and RSS-210 of Industry
Canada.
Operation of this equipment requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry Service.
UL/IEC/EN 60601-1
CAN/CSA C22.2 No.601.1
IEC/EN 60601-1-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4
IEC/EN 60601-2-27
IEC/EN 60601-2-49
CE marked to the Medical Devices Directive 93/42/EEC
608.025 to 613.975 MHz - FCC Part 95 (ApexPro CH)
ApexPro™
± 5% @ 15 Hz
100 dB minimum @ 60 Hz
± 5000 VDC, 360 joules into 100 ohm
Defibrillator recovery time recovers within 5 seconds
ApexPro: ± 2 mV to ± 700 mV (RTI); 100 µsec to 2 msec;
either polarity
ApexPro CH: ± 2 mV to ± 700 mV (RTI); 100 µsec to 2 msec;
either polarity; on multiple leads
Meets UL/IEC 60601-1
2 ports at 9600 baud asynchronous
13.7 cm (5.38 in)
7.3 cm (2.875 in)
2.3 cm (0.91 in)
141.8 g (0.275 lb) without battery; 170.1 g (0.375 lb) with
battery
2001989-200E