Viewing real-time patient data
Placement in the blood pressure cuff
Safety considerations
7-32
Microphone Cable
K-sound Microphone
Microphone Pad
As already stated, using the microphone pad is recommended, especially in the case
of ambulatory patients or patients with weak K-sounds. However, as an alternative,
the microphone can also be placed directly in the blood pressure cuff. Follow the
directions below.
NOTE
Blood pressure readings taken with the microphone in the blood pressure cuff
may not be as accurate as readings obtained when using the microphone pad.
1.
Remove the bladder from the cuff.
2.
Turn the cuff bladder pouch inside out to expose the microphone pocket.
3.
Open the Velcro pocket flap and gently insert the microphone into the pocket.
4.
When the microphone is completely inserted, close the Velcro flap over the
microphone cable and turn the cuff right side out.
5.
Replace the bladder and exit the bladder hose and microphone cable out of the
same exit site, either right arm or left arm, as marked on the cuff.
WARNING
The Accutracker DX blood pressure monitor is designed for use
with adult patients only. Do not use on neonates or on patients
known to be susceptible to bruising.
Do not attach the blood pressure cuff to a limb being used for IV
infusions as the cuff inflation can block the infusion, causing harm
to the patient.