GE ApexPro Operator's Manual page 120

Telemetry system
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Viewing real-time patient data
7-12
4.
Change any of the undimmed setting options. When an option appears dimmed,
you cannot change it unless you enter the service-level password.
Option
Display Lead
Relearn button
Size
Detect Pace
Pace Help button
Lead Analysis
ApexPro™
ECG control settings
Function
Set the top or first lead displayed on the monitor and on the
CIC Pro center. This is the lead data that prints during an
alarm or manual graph.
Relearn the selected patient's ECG rhythm after changes
occur to heart rate or rhythm. The CIC Pro center uses 14
current complexes to relearn the patient's ECG pattern.
Remove the V Fail message or activate the second V-lead
when changing between a 5-and 6-leadwire set on the
ApexPro CH transmitter.
NOTE
The heart rate value appears briefly as Xs during the
relearn process and returns to numerics when the relearn
is complete.
Set the waveform size. 1X is standard.
Set the pacemaker detection mode:
Pace 1: Alternate pacemaker detection mode when Pace
2 does not adequately detect pacemaker spikes.
Pace 2: Normal pacemaker detection mode.
For more information, refer to
Monitoring pacemaker patients
on page
7-13.
View solutions to common pacemaker detection problems.
Set the leads for ECG and arrhythmia data processing:
Single-Lead: Use the top Display Lead.
Multi-Lead: Use leads I, II, III and V lead.
063A
2001989-200E