Connecting the drain hose

- A non-return valve is incorporated into the drain system in the ma-
chine to prevent drainage water flowing back into the machine via
the drain hose.
- The machine drainage hose should be connected to a separate
drain for the machine only. If no separate drain is available, we re-
commend connecting it to a dual-chamber siphon.
- The on-site connection point, measured from the lower edge of
the machine, should be positioned at a height between 0.3 m and
1.0 m. If it is lower than 0.3 m, the drain hose must be laid in a coil
at a height of at least 0.3 m.
- The drainage system must be able to accommodate a minimum
drainage flow of 16 l/min.
- The drainage hose is approx. 1.4 m long and flexible with an in-
ternal diameter of 22 mm. Hose clips for the connection are sup-
plied.
- The drain hose must not be shortened.
- The drain hose can be extended to 4.0 m in length using a connec-
tion piece to attach a further length of hose. The drainage length
must not be longer than 4.0 m.
See installation diagram supplied.