GE USM 36 Series Technical Reference And Operating Manual page 255

Ultrasonic flaw detector
Table of Contents

Advertisement

Function directory
Function
Function group
OUTPUT SELECT
CONFIG2
FINISH
CNDAC
FINISH
DAC/TCG
FINISH
JISDAC
RECORD
AUTOCAL
RECORD
CNDAC
RECORD
DAC/TCG
RECORD
JISDAC
ENTER
FILES
AUTO80
CNDAC
AUTO80
DAC/TCG
AUTO80
JISDAC
B REFERENCE
AWS D1.1
B START MODE.
CONFIG2
B-START
GATE B
USM 36
Operating level Description
second
first
Finishing of the recording of points (curve points)
for the DAC
first
Finishing of the recording of points (curve points)
for the DAC
first
Finishing of the recording of points (curve points)
for the DAC
first
Start of the semi-automatic calibration function
first
Recording of points (curve points) for the DAC
first
Recording of points (curve points) for the DAC
first
Recording of points (curve points) for the DAC
second
Execution of the file processing mode selected using the
function ACTION
first
first
first
first
second
Basis for the starting point of the gate B
first
Starting point of the gate B
Issue 2 (12/2013)
9 Appendix
see page
5-76
5-111
5-93
5-104
5-28
5-111
5-93
5-104
6-2
5-111
5-93
5-104
5-40
5-26
5-21
9-3

Hide quick links: