Probe Data - GE USM 36 Series Technical Reference And Operating Manual

Ultrasonic flaw detector
Table of Contents

Advertisement

5 Operation

Probe data

#
Probe name
1
B1-S
2
B2-S
3
B4-S
4
MB2-S
5
MB4-S
6
MB5-S
7 ... 9
MWB ...-2
10 ... 12
MWB ...-4
13 ... 15
SWB ...-2
16 ... 18
SWB ...-5
19 ... 21
WB ...-1
22 ... 24
WB ...-2
5-132
Wavelength
Minimum diameter
in steel
of side-drilled hole
[mm]
[mm]
6.0
9.0
3.0
4.5
1.5
2.3
3.0
4.5
1.5
2.3
1.2
1.8
1.6
2.4
0.8
1.2
1.6
2.4
0.7
1.1
3.3
5.0
1.6
2.4
Issue 2 (12/2013)
Evaluation according to the DGS method
Near-field length
Min. distance
in steel
in steel
[mm]
[mm]
23
35
45
68
90
135
8
12
15
23
20
30
15
23
30
45
39
59
98
147
45
68
90
135
USM 36