GE USM 36 Series Technical Reference And Operating Manual page 153

Ultrasonic flaw detector
Table of Contents

Advertisement

General setup
The following settings are available:
● A (+)
The alarm is triggered if the gate A is exceeded.
● B (+)
The alarm is triggered if the gate B is exceeded.
● A / B (+)
The alarm is triggered if one of the gates A or B is ex-
ceeded.
● A (-)
The alarm is triggered if the gate A is not reached.
● B (-)
The alarm is triggered if the gate B is not reached.
● A / B (-)
The alarm is triggered if the two gates A and B are not
reached.
USM 36
– Switch to the second operating level.
– Select the function group CONFIG2.
– Select the function OUTPUT SELECT.
– Press one of the selector keys to choose the required
setting.
Issue 2 (12/2013)
5 Operation
5-77