Guidance and manufacturer's declaration – electromagnetic immunity
The NU-IP40S is intended for use in the electromagnetic environment specified below. The customer or the user of the NU-IP40S
should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips, short
< 5%
interruptions and
(> 95% dip in
voltage variations
for 0.5 cycle
on power supply
input lines
40%
(60% dip in
IEC 61000-4-11
for 5 cycles
70%
(30% dip in
for 25 cycles
< 5%
(> 95% dip in
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE:
is the a.c. mains voltage prior to application of the test level.
T
Compliance level
±6 kV contact
±8 kV air
±1 kV for input/
output lines
Not applicable
Not applicable
T
)
T
T
)
T
T
)
T
T
)
T
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the NU-IP40S requires
continued operation during power mains interruptions, it is
recommended that the NU-IP40S be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital
environment.
5