GE Panda iRes Warmer Operation And Maintenance Manual page 88

Table of Contents

Advertisement

Patient Temperature Control Mode
In baby mode, the system performs the following adjustment every 30 seconds. The Baby Condition
column indicates the difference between the baby skin temperature and the baby mode set
temperature.
Baby Condition
Warm by 0.15°C or more
Warm by 0.06°C to 0.14°C
Warm by 0.05°C or less
Cold by 0.05°C or less
Cold by 0.06°C to 0.14°C
Cold by 0.15°C to 0.24°C
Cold by 0.25°C to 0.34°C
Cold by 0.35°C or more
Specifications
The product is designed to meet a life span of 8 years. However, with proper maintenance and
repairs, the service life can be extended as long as service parts are available.
The maximum patient weight is 40 kg (88 lbs).
During the seven minute warmup period, the radiant heater is at 100% power (~27 mW/cm
the radiant heater has been running in warmup mode for seven minutes, the heater power setting
drops to 50% (10 to 15 mW/cm
E-2
Panda iRes Warmer
).
2
M1110737 006
Heater Output Adjustment
Decrease by 5%
Decrease by 1%
No adjustment
No adjustment
Increase by 1%
Increase by 5%
Increase by 10%
Increase by 15%
© 2007 General Electric Company, all rights reserved.
). After
2