GE Panda iRes Warmer Operation And Maintenance Manual page 50

Table of Contents

Advertisement

Attaching the Skin Temperature Probe (continued)
Illustration
NOTE:
The Panda iRes Warmer cannot differentiate between an increase in core temperature with
cold skin (fever), and low core and skin temperatures (hypothermia). Patient temperature
should be verified with an axillary thermometer.
3-10
Panda iRes Warmer
Step
2
To attach the disposable probe (single use
only), peel the paper backing from the adhe-
sive side and apply to the baby's skin, with
the reflective foil side up. To attach the reus-
able probe, use the heat reflecting patch.
Place the metal side of the probe against
the baby's skin, peel the paper backing from
the adhesive side of the patch and place the
patch over the probe with the reflective foil
side up.
3
Route the probe wire through the gaps at
the corner of the bedside panels or through
the one of the slots in the rear bedside panel.
4
Plug the temperature probe into the jack.
M1110737 006
Description
© 2007 General Electric Company, all rights reserved

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Giraffe

Table of Contents