3. Loosen the three screws and remove the shield plate for signal cables (see the figure
below).
4. Pull out the connectors for the following cables depending on the installation situation:
– Power cable
– 26-pin ribbon cable to display
– 12-pin ribbon cable to analyzer modules AM and option module OM 2.2
– 26-pin ribbon cable to the analyzer modules
– Connecting cable to the terminal block (TB)
– If applicable, Ethernet-cable on the processing unit.
Performing installation
Option module OM 2.1
Screw, printed circuit board holder
Printed circuit board holder
Screw, electronics base plate
Figure 7-10
Installing option module, example OM 2.1
1. Loosen the screws
2. Loosen the screw
3. Remove the blanking cover plate for option modules.
Wall-mounted device
Operating Instructions, 05/2018, A5E31930403-06
Installing / removing and connecting analyzer and option modules
Screws option module (2 units)
Electronics base plate
Screw, electronics base plate
Shielding plate signal lines
and
. Remove the electronics base plate
and remove the printed circuit board holder
7.7 Installing option modules
⑪⑫
⑦⑧⑨
.
.
71