Maintenance and servicing

11.10 Eliminating faults and errors

11.10
Eliminating faults and errors
Table 11- 1
Error messages
Fault message
Description
Invalid device con-
No analyzer module was detected. At least one analyzer module is
figuration
required to use the device.
Firmware versions
The firmware of the basic device and the analyzer modules must be
incompatible
compatible with one another.
Unknown module
An analyzer module has been detected which is not supported in
type detected
the firmware version.
Memory error de-
A hardware fault, faulty firmware or no firmware at all has been
tected
detected in the following analyzer modules, e.g., AM1. Possible
causes:
In this case, the analyzer module starts in bootloader mode and is
ready to perform a firmware update.
If the error still exists despite the firmware update, the analyzer
module PCB must be replaced.
Cable fault detected
The installed analyzer modules could not be recognized correctly.
Possible causes: Faulty or incorrectly connected CAN bus cable.
136
No firmware is present on the analyzer module.
The existing firmware is invalid.
The program memory is faulty.
Measures
Install at least one analyzer
module.
Please contact the service
department.
Please contact the service
department.
Please contact the service
department.
Check the CAN bus connec-
tions.
Wall-mounted device
Operating Instructions, 05/2018,
A5E31930403-06