4.5 Analog Voltage Output Scale

Define an analog voltage output range in proportion to the detected value.
The full-scale is set in the range from 0 to 5.0V before shipment from the fac-
tory.
An analog voltage output range, inclination and other features of the detect-
ed value can be entered.
The setting method includes three variations, similar to the display scale:
one-point scale, two-point scale, and INV.
Or select the default setting to return to the factory shipment state.
Only the voltage output can be changed with this function. To change the
displayed value according to the detected value, use the display scale.
Notice
However, if voltage (volts) is selected as a display unit, the indication
changes after analog voltage output scaling.

4.5.1 One-Point Scaling

Only the offset changes while the voltage output span (inclination) remains
unchanged. With one-point scaling, a "SLOPE" function is available with
which the inclination reverses around the point designated for scaling.
Change "0.5mm 1.2500" to "0.5mm 1.5000".
Voltage
(V)
5.0000
1.2500
0
0.5mm
2mm
Voltage
(V)
5.2500
5.0000
1.5000
0.2500
0
0.5mm
2mm
Change the inclination with "SLOPE".
Voltage
(V)
5.2500
1.5000
0.2500
-2.2500
0.5mm
2mm
81