Troubleshooting the MSD
3

General symptoms

General symptoms

This section describes symptoms you might observe when first turning on the
GC/MSD system. All of these symptoms would prevent operation of the system.

GC does not turn on

This refers to a condition in which nothing happens when the GC is switched on.
The GC fans do not turn on and the keypad display does not light.
• Disconnected GC power cord
• No voltage or incorrect voltage at the electrical outlet
• Failed fuse in the GC
• GC power supply is not working correctly

MSD does not turn on

This refers to a condition in which nothing happens when the MSD is switched on.
The foreline pump does not start. The cooling fan for the high vacuum pump does
not turn on. The status LEDs on the front of the MSD do not light.
• Disconnected MSD power cord
• No voltage or incorrect voltage at the electrical outlet
• Failed primary fuses
• MSD electronics are not working correctly

Foreline pump is not operating

This refers to a condition where the MSD is receiving power (the fan is operating
and the power LED on the status display is lit) but the foreline pump is not
operating.
• Disconnected foreline pump power cord
• Malfunctioning foreline pump