To Remove The Diffusion Pump - HP 5973 Hardware Manual

Mass selective detector
Hide thumbs Also See for 5973:
Table of Contents

Advertisement

4

To remove the diffusion pump

To remove the diffusion pump
Materials needed:
Aluminum foil, clean
Gloves, oil-resistant
1 Vent the MSD (page 52).
2 Separate the MSD from the GC (page 112).
: $ 5 1 , 1 *
7KH GLIIXVLRQ SXPS RSHUDWHV DW YHU\ KLJK WHPSHUDWXUHV 0DGH VXUH LW KDV FRROHG EHIRUH
KDQGOLQJ LW
3 Disconnect the foreline gauge assembly from the diffusion pump outlet.
The foreline gauge cable can be disconnected or can remain connected to the fore-
line gauge. Place the O-ring assembly on clean aluminum foil.
4 Disconnect the diffusion pump temperature sensor wires from the wiring
harness.
These are on the side of the diffusion pump not shown in the illustration.
5 Disconnect high vacuum power (
the MSD.
This is the thick black cable that emerges near the bottom of the pump.
6 Support the diffusion pump with one hand.
7 Remove the KF50 clamp.
8 Lower the diffusion pump.
9 Remove the O-ring assembly from the top of the diffusion pump.
The O-ring will have diffusion pump fluid on it and will be very sticky. Place the
O-ring on clean aluminum foil to avoid contaminating your lab bench and to avoid
having the O-ring pick up dust and dirt.
114
Maintaining the MSD
HIVAC POWER
) cable from the back panel of