To connect the optional gauge controller

To connect the optional gauge controller

Materials needed:
Gauge controller (HP 59864B)
Power cord
Triode gauge cable (8120-6573)
: $ 5 1 , 1 *
1HYHU FRQQHFW RU GLVFRQQHFW WKH FDEOH IURP WKH WULRGH JDXJH WXEH ZKLOH WKH 06' LV XQGHU
YDFXXP 5LVN RI LPSORVLRQ DQG LQMXU\ GXH WR EURNHQ JODVV H[LVWV
& $ 8 7 , 2 1
Do
model can be used for diagnostic purposes RQO\.
& $ 8 7 , 2 1
Be sure to orient the cable and the gauge tube as illustrated. Excessive force on the pins
can break the tube.
1 Connect the triode gauge cable to the triode gauge tube.
2 Connect the other end of the triode gauge cable to the gauge controller.
3 Connect the power cord to the gauge controller.
4 Connect the other end of the power cord to an appropriate electrical
outlet.
If you wish to share one controller among MSDs, obtain one cable for each instru-
ment. Leave a cable connected to the triode gauge tube on each MSD. This will
avoid having to vent the MSD before connecting the controller.
60
Operating the MSD
QRW
use an HP 59864A (older model) triode gauge controller during data acquisition. This