4.3.17 Preset Buttons

By registering magnifications that is frequently used (two magnification levels can be
memorized within the range of 15x to 10,000x),your target magnification can be set with a
simple click of the [Preset] button. For a description of how to set a Preset magnification, see
4.3.33 Preset Magnification Window.
NOTE : If presetting two magnifications levels with a large magnification difference, for
example, a low magnification (e.g. 100x or lower) and a high magnification (e.g.
10,000x), an deviation of the field of view can occur. In such a case, change
magnifications by using the [Magnification] button. It is recommended that the
preset magnification should be set within a magnification of 5,000x or lower.
NOTE : If the magnification set as a preset value is not allowable in consideration of the
current observation condition and the value of D (i.e. the magnification is 15x, 18x,
20x, 25x or 30x), the magnification will not be switched, even though the [Preset]
button is clicked. (See Table 4.3.16.)
Figure 4.3.17 Preset Buttons