3. Loop at Fault
Location in Matrix
Description of Function
VH 07
If any fault malfunction occurs, the loop can be set to either "-
10%" (=3.6 mA), to "110%" (=22 mA) or "Hold" (stays at last
valid value).
4. Selecting the Analogue Output Offset (4 mA)
Location in Matrix
Description of Function
VH 10
The dewpoint value corresponding to the analogue output
offset (4 mA) is entered here.
Caution! Ensure this is always at least 20 OC below the value
assigned to 20 mA.
Default: -90 oC
5. Selecting the Analogue Output Span (20 mA)
Location in Matrix
Description of Function
VH 11
The dewpoint value corresponding to the analogue output span
(20 mA) is entered here.
Caution! Ensure this is always at least 20 oC above the value
assigned to 4 mA.
Default: + 10 oC
6. Setting the Span Value for the ppm-v Range
Location in Matrix
Description of Function
VH 12
Selection of this field sets the span value for the ppm-v range.
NOTE: The offset is always 0 ppm-v. Do not exceed 1000
ppm-v
Default: 100