Description Of The Dewpro Mmr 31 Programming Matrix - GE DewPro MMY31 Installation And Operation Manual

Trace moisture transmitter
Hide thumbs Also See for DewPro MMY31:
Table of Contents

Advertisement

DewPro MMY31
4.2 Description of the DewPro MMY31 Programming Matrix
In the DewPro mid-range moisture transmitter with display option, a matrix-style input is
used for programming the unit of measure, measuring range, error status of output, and
output adjustment. For users of other General Eastern equipment, this 'GEI matrix" format is
familiar. The following describes the features and usage of the various matrix location as
they apply to the MMY 31.
Fig. 19
The display of the DewPro MMY 31 continuously shows the current matrix location using
the vertical (V) and horizontal (H) coordinates to designate the row and column,
respectively. The bar graph represents the output current in an analogue fashion (refer to
Figure 19). See the Appendix for an enlarged overview of the matrix.
MMY 31
Moisture Unit
Loop Range
Constant;
Loop
Hardware
Calibration
Access Key
Misc. Setup
Fig. 20
Movement through the matrix is accomplished by using the "V" and "H" buttons to move
to another row or column. For example, beginning at VH 00 and successively pressing
"V", leads the user to VH 10, VH 20, VH 30, VH 40, VH 50, VH 60, VH 70, VH 80, VH 90
and back to VH 00. At any location where a value may be changed by the user, the
desired value is programmed using the "+" and "-" buttons.
GE General Eastern
8888
----------
Numeric
-
Display
Bar Graph
H0
H1
H2
Select
Display
Moisture Unit
V0
Moisture
see table
Value
below
Dewpoint °C
Dewpoint °C
ppm
v
V1
4 mA
20 mA
mA
V2
Pressure
Constant for
V3
ppmv comp.
(bar)
V4
V5
V6
V7
V8
Display
Display
V9
Unit ID
Present Error
Previous Error
Installation and Operation Manual
VH
VH Position
88
Indicators
User Interface
Keys
H3
H4
H5
H6
Reset To
Software
Defaults
Version
50 = Reset
H7
H8
H9
Loop 1 at Fault
Loop 1 Raw
0= -10%,
Reading
1=110%, 2=Hold
Loop 1 D/A
Loop 1 D/A
Calibration
Calibration
Low
High
Input Locking
50 = Unlock
System Reset
50 = Reset
11

Advertisement

Table of Contents
loading