Code
Error Message
1690
The instruction manual does not
exist.
Install the instruction manual.
The instruction manual cannot
1691
be displayed.
Confirm if the PDF file browser
has been installed.
1706
Observation screen size is
changed.
Do you save this setting?
The input value of the
1707
observation screen size is
unacceptable.
This size returns to the previous
registered value.
Re-starting up the program is
1720
required for changing the
language.
Do you exit the application?
Re-starting up the program is
1721
required for changing the option
setting.
Do you exit the application?
It is time to maintain the Turbo
1735
Molecular Pump.
It is recommended to replace it
with a new one.
Countermeasure
This message appears if the TM3000 manual (in
PDF format) is not installed on the PC,
preventing the display of the manual.
Install the TM3000 manual from the supplied
Manual CD.
This message appears if the PDF file read
software (Adobe Acrobat Reader) not installed
on the PC, preventing the display of the manual.
Install the PDF file read software (Adobe Acrobat
Reader) from the supplied Manual CD.
This confirmation message appears when the
[Close] button is clicked without clicking the
[SET] button after changing the observation
screen size.
Clicking the [OK] button changes the size and
closes the [Setup] screen. Clicking the [Cancel]
button closes the [Setup] screen without
changing the size.
This message appears when the observation
screen size that was entered is out of range.
Clicking the [OK] button resets the observation
screen size to the currently stored settings.
This message appears when the modification of
display language is selected.
Clicking the [OK] button applies the change in
display language and starts the termination
processing. Clicking the [Cancel] button cancels
the modification of display language.
This message appears when option settings are
modified.
Clicking the [OK] button applied the change in
option settings and starts the termination
processing. Clicking the [Cancel] cancels the
modification of option settings.
This message appears when the turbo molecular
pump run time has reached a specified limit (720
hours).
Click the [OK] button to dismiss the message.
5 - 45