1.3 Safety Information
Danger
Risk to persons —
– The equipment is not designed for use in areas
where an explosion hazard may occur. Explosion
hazards may result from the use of flammable
anesthetic mixtures with air or with oxygen,
nitrous oxide, skin cleansing agents or
disinfectants.
8
Application, Safety Information
Risk to persons —
– Equipment may be connected to other
equipment or to parts of systems only when it
has been made certain that there is no danger to
the patient, the operator, or the environment as
a result. In those instances where there is any
element of doubt concerning the safety of
connected equipment, the user must contact the
manufacturers concerned or other informed
experts as to whether there is any possible
danger to the patient, the operator, or the
environment as a result of the proposed
combination of equipment. Compliance with the
standard IEC 60601-1 must always be ensured.
– TONOPORT V may be connected to CASE™ or to
a PC with the CardioSoft program. While
connected to any of these devices, TONOPORT V
must be disconnected from the patient.
– Chemicals required for the maintenance of the
equipment, for instance, must under all
circumstances be prepared, stored, and kept at
hand in their specific containers. Failure to
observe this instruction may have severe
consequences for the patient.
– The equipment has no protection against the
ingress of liquids. Liquids must not enter the
equipment. Equipment into which liquids have
entered must be inspected by a service
technician before use.
– Before cleaning, TONOPORT V must be
disconnected from other equipment (CASE™,
PC).
– Dispose of the packaging material, observing
the applicable waste-control regulations. Keep
the packaging material out of children's reach.
TONOPORT V
Warning
2001589-085 Revision G