Guidance and Manufacturer's Declaration – Electromagnetic Immunity
TONOPORT V is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that TONOPORT V is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD) to EN 61000-4-2/
IEC 61000-4-2
Electrical fast transient/
burst to EN 61000-4-4/
IEC 61000-4-4
Surge to EN 61000-4-5/
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines to EN
61000-4-11/IEC61000-4-
11
Power frequency (50/
60 Hz) magnetic field to
EN 61000-4-8/IEC 61000-
4-8
NOTE
U
is the AC mains voltage prior to application of the test level.
T
31
Electromagnetic Compatibility (EMC)
EN/IEC 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
(> 95 % dip in U
T
for 0.5 cycles
40 % U
(60 % dip in U
) for
T
T
5 cycles
70 % U
(30 % dip in U
) for
T
T
25 cycles
< 5 % U
(> 95 % dip in U
T
for 5 s
3 A/m
TONOPORT V
Compliance
Electromagnetic Environment -
Level
± 6 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
± 8 kV
synthetic material, the relative humidity
should be at least 30%.
not
Mains power should be that of a typical
applicable
commercial or hospital environment.
not applicable
not applicable
Mains power should be that of a typical
commercial or hospital environment.
not applicable
)
not applicable
Mains power should be that of a typical
T
commercial or hospital environment. If
the user of TONOPORT V requires
not applicable
continued operation during power mains
interruptions, it is recommended that
TONOPORT V be powered from an
not applicable
uninterruptible power supply or a battery.
)
T
not applicable
3 A/m
Power frequency magnetic fields should
be at levels characteristics of a typical
location in a typical commercial or
hospital environment.
Guidance
2001589-085 Revision G