Guidance and Manufacturer's Declaration – Electromagnetic Immunity
TONOPORT V is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that TONOPORT V is used in such an environment.
Immunity Test
Conducted RF to
EN 61000-4-6/IEC 61000-
4-6
Radiated RF to EN 61000-
4-3/IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
2001589-085 Revision G
Electromagnetic Compatibility (EMC)
EN/IEC 60601 Test Level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
TONOPORT V
Compliance
Electromagnetic Environment -
Level
Portable and mobile RF communications
equipment should be used no closer to any
part of TONOPORT V, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
3 Vrms
d = 1.17
d = 1.17
3 V/m
d = 2.33
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey
, should be less than the
compliance level in each frequency range
Interference may occur in the vicinity of
equipment marked with the following
symbol
Guidance
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz
P
b
.
32