Master Reset (P999)

P999
MASTER RESET
Global Master Reset
A
performed:
When performing a Master Reset, the bar graph will indicate progress
When performing a Master Reset, the bar graph will indicate progress
When performing a Master Reset, the bar graph will indicate progress
When performing a Master Reset, the bar graph will indicate progress
Following a Global Master Reset, complete reprogramming is required.
Following a Global Master Reset, complete reprogramming is required.
Following a Global Master Reset, complete reprogramming is required.
Following a Global Master Reset, complete reprogramming is required.
Point Number Master Reset
A
Point Number to preset values) should be performed if a Point Number is being
reassigned to a different vessel.
To perform a Global Master Reset:
To perform a Point Number Master Reset:
7ML19985AN02
(reset all parameters to original values) should be
prior to initial programming if arbitrary Parameter Values were
used during a "bench test".
after replacing the SITRANS LU10 EPROM with a different
software revision number.
1. With the Parameter Number field accessed, key in 999.
2. With the Point Number field accessed, key in 00.
3. Press
, "C.ALL" is displayed until the reset is complete.
1. With the Parameter Number field accessed, key in 999.
2. With the Point Number field accessed, key in the Point Number to be
reset (1 - 10).
3. Press
, "C.ALL" is displayed until the reset is complete.
P999 is not accessible via SmartLinx Interface.
P999 is not accessible via SmartLinx Interface.
P999 is not accessible via SmartLinx Interface.
P999 is not accessible via SmartLinx Interface.
SITRANS LU10 – INSTRUCTION MANUAL
from 0 to 100%.
from 0 to 100%.
from 0 to 100%.
from 0 to 100%.
(reset the non-global parameters of the displayed
Page 77