can use USB to connect the 405 TS to the computer or the RS232 serial port to connect
to a BioStack or similar robotic device. But you cannot use both ports
simultaneously, i.e. make sure only one cable is plugged in at a time.

Washer Problems

Review these suggestions to correct problems your washer is having:
Vacuum Pump Problems on page 102
l
Fluid Aspiration on the next page
l
Fluid Delivery on page 98
l
Fluid Leakage on page 100
l
Washer Manifold Movement on page 101
l
Microplate Carrier Movement
Possible
Problem
Cause
Aspiration
Microplate not
tubes not
properly
entering
seated or
wells
strips not
correctly.
level.
Aspirate tubes
position is too
wide for a
movement.
Aspirate tubes
bent.
Loud,
Plate carrier is
annoying
rubbing
noise during
against glide
operation.
strip.
384-well
Microplate
plates not
carrier not
properly
moving
washed.
correctly in the
Y-axis.
What To Do
Reseat microplate in carrier or strips in holder.
Make sure the carrier is clean.
Try a different microplate or strip holder. If the
problem is unresolved, the carrier may have to be
realigned. Contact BioTek TAC.
Change the horizontal, X- or Y-axis aspirate position
in the protocol.
Contact BioTek TAC.
Thoroughly clean the microplate carrier and exterior
surface as recommended in the maintenance
procedure.
Clean the plate carrier system. If problems persist,
contact TAC for assistance cleaning the carrier
transport arm.

Washer Problems| 95

BioTek Instruments, Inc.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 405 TS and is the answer not in the manual?

Questions and answers

Julia Driscoll
February 14, 2025

Cannot locate the Monthly Maintenance for the Vacuum Filtration Washer System on the BioTek 405TS Microplate Washer

Table of Contents

Save PDF