Page 4
BioTek is a registered trademark, and Epoch™, Gen5™, BioStack™, and Take3™ ® and Take3 Trio™ Micro-Volumne Plate are trademarks of BioTek Instruments, Inc. BioCell™ is a trademark of BioTek Instruments and is patented under U.S. patent number 5,963,318. ...
Winooski, Vermont 05404-0998 USA Global Service and Support BioTek instrument service and repair is available worldwide at one of BioTek’s International Service Centers and in the field at your location. To arrange for service or repair of your instrument, contact the office nearest you; visit www.biotek.com...
Chapter 1, Introduction: Updated wording for getting the shipping address in Product Support & Services Chapter 3, Control Using Gen5: Included instructions for setting up the Epoch as a Plug & Play reader Chapter 4, Qualification: Removed mention of Software Documentation, Wavelength Table Verification, and “Run Assay”...
Page 9
“Decontamination” section text to match other instruments’ decontamination procedure. Chapter 5, Qualification: Added 340 nm Absorbance Test Plate PN 7260551 to the “Recommended Qualification Schedule” section; updated text of System Test and Liquid Test sections to match other instruments’ Qualification chapters. Epoch Operator’s Manual...
Caution indicates potential damage to the instrument and tells Caution you how to avoid the problem. text is primarily used for emphasis. Note Bold This icon calls attention to important information. BioTek Instruments, Inc.
Warranty and Product Registration Please take a moment to review the warranty information that shipped with your product. Please also register your product with BioTek to ensure that you receive important information and updates about the product(s) you have purchased.
If a spill occurs while a program is running, abort the program and turn off the instrument. Wipe up all spills immediately. Do not operate the instrument if internal components have been exposed to fluid. Call BioTek TAC for assistance.
Precautions The following precautions are provided to help avoid damage to the instrument: The instrument should be serviced by BioTek-authorized Caution: Service. service personnel. Only approved spare parts should be used for Caution: Spare Parts.
Immunity. Verification of compliance was conducted to the limits and methods of the following: EN 61000-4-2, Electrostatic Discharge EN 61000-4-3, Radiated EM Fields EN 61000-4-4, Electrical Fast Transient/Burst EN 61000-4-5, Surge Immunity EN 61000-4-6, Conducted Disturbances from RFI EN 61000-4-11, Voltage Dips, Short Interruptions and Variations BioTek Instruments, Inc.
In order to maintain compliance with FCC regulations, shielded cables must be used with this equipment. Operation with non-approved equipment or unshielded cables is likely to result in interference to radio and television reception. Epoch Operator’s Manual...
Part 1: general requirements” Canadian Standards Association CAN/CSA C22.2 No. 61010-1 • “Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use; Part 1: General Requirements” EN 61010 standards, see CE Mark, starting on page xii • BioTek Instruments, Inc.
Attention, risque d’écrasement et électrique pincement Gefährliche elektrische schlag Warnen, Gefahr des Zerquetschens und Klemmen Precaución, riesgo de sacudida eléctrica Precaución, riesgo del machacamiento y sejeción Attenzione, rischio di scossa elettrica Attenzione, rischio di schiacciare ed intrappolarsi Epoch Operator’s Manual...
Page 18
Recogida selectiva de aparatos eléctricos y electrónicos Dispositivo medico diagnostico in vitro Raccolta separata delle apparecchiature elettriche ed elettroniche Consult instructions for use Consulter la notice d’emploi Gebrauchsanweisung beachten Consultar las instrucciones de Consultare le istruzioni per BioTek Instruments, Inc.
Chapter 1 Introduction This chapter introduces the Epoch, describes its hardware and software features, and provides contact information for technical assistance. Product Description ............2 Package Contents ............2 Optional Accessories ............3 Product Support and Service ........... 4 Technical Assistance Center (TAC) ........4...
BioTek’s patented BioCell quartz vessel for 1-cm measurements, and the BioTek Take3 and Take3 Trio Micro-Volume Plate. Two reading speeds: normal and sweep. • Some models of the Epoch are robot-accessible and compatible with the • BioStack Microplate Stacker. Package Contents ...
Take3 Trio Micro-Volume Plate TRIO Some models of the Epoch are compatible with the BioStack Microplate Stacker. The BioStack rapidly and systematically transfers a “stack” of microplates to and from the Epoch’s microplate carrier. Contact BioTek or visit our website to learn more (www.biotek.com).
(see also Appendix C, Error Codes). If you need to return an instrument to BioTek for service or repair, please contact the TAC for a Service Call Notice (SCN) number and the shipping address.
Chapter 2 I nstallation This chapter includes instructions for unpacking and setting up the Epoch, connecting to a PC, and repackaging the instrument. Product Registration ............6 1: Unpack and Inspect the Instrument ......6 2: Select an Appropriate Location ........6 3: Remove the Shipping Hardware ........
2: Select an Appropriate Location Install the Epoch on a level, stable surface in an area where ambient temperatures between 18°C (64°F) and 40°C (104°F) can be maintained. The reader is sensitive to extreme environmental conditions. Conditions to avoid are: •...
Excessive vibration: machines that cause the surface to vibrate. If you are installing the BioStack for operation with the Epoch, you may wish to seat the instruments in their aligning plates now. Refer to the BioStack Operator’s Manual for more information.
Connect the power supply’s outlet plug to the 24-VDC connector on the rear of the instrument. Tighten the plug barrel to retain the plug. Plug the other end of the power cord into an appropriate power receptacle. BioTek Instruments, Inc.
If the test is successful, the reader is ready for use. • If the test fails, note any error codes that are displayed in Gen5, and contact • BioTek. Appendix B, Error Codes, contains a list of error codes that may appear in Gen5. Epoch Operator’s Manual...
Select Plug & Play An Epoch must be connected via USB to the computer and turned on to appear in the Available Plug & Play Readers list. Set the Com Port to the computer’s COM port to which the reader is •...
If you remain unable to get Gen5 and the reader to communicate with each other, contact BioTek’s Technical Assistance Center. Operational/Performance Qualification Your Epoch was fully tested at BioTek prior to shipment and should operate properly following the successful completion of the installation and setup procedures described throughout this chapter.
• rubber plug (PN 19610) If you need to ship the Epoch to BioTek for service or repair, be sure to use the original packaging materials. Other forms of commercially available packaging are not recommended and can void the warranty.
BioTek’s Technical Assistance Center for guidance. Ensure that the Epoch carrier shipping hardare has been attached to the reader’s carrier as instructed in the preceding section, A ttach the Shipping Hardware...
Page 32
| Chapter 2 Installation Figure 4: Packing the Epoch BioTek Instruments, Inc.
Chapter 3 Getting Started The Epoch can be controlled only using software installed on a PC connected to the reader via the computer's USB port. This chapter describes how to configure BioTek Gen5 to control the reader. Gen5 Software .............. 16 Protocols and Experiments .........
| Chapter 3: Getting Started Gen5 Software BioTek Gen5 software supports all Epoch reader models. This section provides brief instructions for working with Gen5 to create protocols and experiments. Refer to the Gen5 Help system for more information. Protocols and Experiments In Gen5, a protocol contains instructions for controlling the reader and (optionally) instructions for analyzing the data retrieved from the reader.
• Although the Epoch supports standard flat, U-bottom, and V-bottom microplates, the reader achieves optimum performance with flat-bottomed wells. Non-uniformity in the optical density of the well bottoms can cause loss of •...
Page 36
<3% in long kinetics may also have a destructive effect. If the experiment is incubated, it will accelerate the deterioration of chamber components. When in doubt about the use of acids, corrosives, or solvents; please contact TAC@biotek.com. BioTek Instruments, Inc.
| Chapter 4: Preventive Maintenance Overview A general preventive maintenance (PM) regimen for the Epoch includes periodically cleaning all exposed surfaces and decontaminating the instrument before storage or shipment. This chapter includes instructions for the following: • Routine Cleaning Procedure, page 21 •...
(not decontaminated) with a cloth moistened (not soaked) with water or water and a mild detergent. Procedure Turn on the Epoch and press the carrier eject button to eject the microplate carrier. Turn off and unplug the reader from the power supply.
Wipe the plate carrier and all exposed surfaces of the instrument. If detergent was used, wipe all surfaces with a cloth moistened with water. Use a clean, dry lint-free cloth to dry all wet surfaces. If liquid is spilled inside the reader, call BioTek TAC for cleanup instructions. Decontamination Any laboratory instrument that has been used for research or clinical analysis is considered a biohazard and requires decontamination prior to handling.
The bleach solution is caustic; wear gloves and eye protection when handling the solution. Turn on the Epoch and press the carrier eject button to eject the carrier. Turn off and unplug the reader from the power supply. Prepare an aqueous solution of 0.5% sodium hypochlorite (NaClO, or bleach).
Page 42
| Chapter 4: Preventive Maintenance BioTek Instruments, Inc.
Chapter 5 Instrument Qualification This chapter contains procedure for qualifying the initial and ongoing performance of the Epoch. Overview ..............26 IQ/OQ/PQ ..............26 Recommended Qualification Schedule ....... 27 System Test ..............28 Absorbance Plate Test ............ 31 Define the Absorbance Test Plate Parameters ....31 Running the Absorbance Plate Test ......
Operational Qualification (OQ), and Performance Qualification (PQ) procedures for all models of the Epoch Microplate Reader. Every Epoch reader is fully tested at BioTek prior to shipment and should operate properly upon initial setup. If you suspect that a problem occurred during shipment, if...
Recommended Qualification Schedule The schedule shown below defines the BioTek-recommended intervals for performance testing for an Epoch used for one shift seven days a week. The risk factors associated with your tests may require that the Operational and Performance Qualification procedures be performed more or less frequently than shown below.
| Chapter 5: Instrument Qualification System Test Each time the Epoch is turned on, it automatically performs a series of tests on the reader’s motors, lamp, and various subsystems. This test can take a few minutes to complete. If all...
Page 47
Signal Air: 9479 41320 Wavelength 2: 352nm Cal Data: Rst: 4 Gain: 1.94 Rst/Gain: 2.062 Test Data: Rst: 4 Gain: 2.08 Rst/Gain: 1.921 Channel: Meas Total Air: 8944 45824 Dark Air (On): 5391 4411 Signal Air: 3553 41413 Epoch Operator’s Manual...
Page 48
Delta 2: 10045 - 10034 = Delta 3: 15630 - 15615 = Delta 4: 21851 - 21835 = MONOCHROMATOR A = +0.000000 B = -0.000390 C = -0.161696 SYSTEM TEST PASS 0000 Reviewed/Approved By: _______________________________________ Date: ________________ Figure 6: System Self-Test report BioTek Instruments, Inc.
An alternate method for confirming accuracy, linearity, and repeatability is Liquid Test 2, described on page 41. To run this test, you need the BioTek Absorbance Test Plate (PN 7260522) with its accompanying data sheet. The Absorbance OD Standards section contains NIST-traceable standard OD •...
Page 50
If the C6 filter is Holmium or Erbium glass, the certificate contains two Spectral Bandpass tables. For the Epoch, which has a bandpass wider than 5 mm for wavelength >285 nm and close to 4 nm for 230–285 nm, we recommend using the table.
When the test of completed, the results report appears. Scroll through the report; every result should show “PASS”. When the test is complete, print the results. A sample test report is provided in Figure 7. Epoch Operator’s Manual...
Absorbance Test Plate dialog. To verify wavelength accuracy, the wavelength of the maximum absorbance is compared with the peak wavelength value entered in the software, which comes from the Peak Wavelength Certificate supplied with the Test Plate. Epoch Operator’s Manual...
Page 54
Make sure the Test Plate is within its calibration certification period. If it is out of date, contact BioTek to schedule a recertification. Check the microplate carrier to ensure it is clear of debris. •...
• repeatability, and alignment of the reader by performing Liquid Test 2. BioTek offers a dye solution (PN 7120779, 25 mL; or 7120782, 125 mL) that can • be used in the stock solution formulation for Liquid Tests 1 and 2, or, if you prefer, you may use the dye solution described in Table 2 on page 39.
The stock solution for Liquid Tests 1 and 2 may be formulated from the chemicals listed below, or by diluting a dye solution available from BioTek. See Procedure A or Procedure B outlined below and on the following page for details.
Weigh out 0.092 gram of FD&C No. 5 yellow dye powder into a weigh boat. Rinse the contents into a 1-liter volumetric flask. Add 0.5 mL of Tween 20, or 5 mL of BioTek’s wetting agent. Make up to 1 liter with DI water; cap and shake well.
Page 58
Turnaround position. This will result in 32 mean values. Perform a mathematical comparison of the mean values for each microwell in its Normal and Turnaround positions (A1/H12, A2/H11, B1/G12, B2/G11, and so on). In order to pass this test, the differences in the BioTek Instruments, Inc.
80% in the third tube, all the way to 10% in the tenth tube. • Dilute using the 0.05% solution of deionized water and Tween 20. This solution can also be made by diluting the BioTek wetting agent 200:1. Epoch Operator’s Manual...
Page 60
Perform a regression analysis on the data to determine if there is • adequate linearity. Because it is somewhat difficult to achieve high pipetting accuracy when conducting linear dilutions, an R Square value of at least 0.9900 is considered adequate. BioTek Instruments, Inc.
Page 61
± 1.0% ± 0.010 OD, then the expected range for the mean of the same well in the H12 position is 1.873 to 1.931 OD. (1.902 x 1.0% = 0.019 + 0.010 = 0.029, which is added to and subtracted from 1.902 for the range.) Epoch Operator’s Manual...
Absorbance Test Plate. This test is optional because the reader has good “front end” linearity throughout its wavelength range. BioTek Absorbance Test Plate PN 7260551 is offered as an alternative to conducting Liquid Test 3. Materials ...
Page 63
H4, and A5 to H6). Perform a regression analysis on the data to determine if there is • adequate linearity. The three average Mean OD values are the “Y” values. The solution concentrations are the “X” values (1.00, 0.75, 0.50). Epoch Operator’s Manual...
Page 64
| Chapter 5: Instrument Qualification Because it is somewhat difficult to achieve high pipetting accuracy when conducting linear dilutions, an R Square value of at least 0.9900 is considered adequate. BioTek Instruments, Inc.
Appendix A Specifications This appendix contains BioTek’s published specifications for the Epoch. General Specifications ............48 Read Specifications............... 49 Optical Performance ............. 50...
General Specifications Microplates The Epoch accommodates standard 6-, 12-, 24-, 48-, 96-, and 384-well microplates with 128 x 86 mm geometry, and the BioTek Take3 and Take3 Trio Micro-Volume Plates. Hardware & Environmental Xenon flash light source, 10W maximum average power (not...
630/450 nm Normal 100 msec Normal 0 msec Sweep Kinetic Measurements 96 Well Mode Delay Time 630 nm Normal 100 msec Normal 0 msec Sweep Mode Delay Time 630 nm Normal 100 msec Normal 0 msec Sweep Epoch Operator’s Manual...
2.000–2.500 OD +/-3% +/-0.005 OD 384-well plate, normal read speed, 100-ms delay after plate movement 0.000–1.500 OD +/-1% +/-0.005 OD 1.500–2.000 OD +/-3% +/-0.005 OD 96-well and 384-well plate, sweep read speed 0.000–1.000 OD +/-2% +/-0.010 OD BioTek Instruments, Inc.
Page 69
Optical Performance | Optics λ range 200 to 999 nm λ accuracy ± 2 nm λ repeatability ± 0.2 nm λ bandpass 5 nm Photodiodes (2). Measurements are reference channel-corrected for Detector light source fluctuation. Epoch Operator’s Manual...
Page 70
| Appendix A: Specifications BioTek Instruments, Inc.
Appendix B Error Codes This appendix lists and describes Epoch error codes that may appear in Gen5. If an error is displayed, refer to Product Support & Service in Chapter 1. Overview ..............54 Diagnostics .............. 54 General Errors .............. 55 Assay Errors—For 2Dnn Error Code ......
Error codes marked with a asterisk (*) will stop an assay from running but will continue to allow the Epoch to communicate with the controlling software and run system and calibration tests. When these errors are first encountered, the status LED will flash and the beeper will sound;...
Filter wheel gearing is worn or dirty. • Defective or broken optical sensor or connection. • Defective motor, motor controller PCB, or cable. • 0233 Monochromator not able to locate home. Probable causes: Misaligned flash lamp • Disconnected photodiode cable • Epoch Operator’s Manual...
Page 74
• Defective flash lamp. • • Defective motor/power PCB or connection to it. • Flash lamp is missing flashes or is not flashing. The optic system does not detect the saturation. • Defective monochromator (low probability). • BioTek Instruments, Inc.
Page 75
There may be an ambient light leak. Ensure that the • plate carrier door is properly closed. Detector PCB or connection failure. • Internal electronic noise may be caused by a faulty • analog PCB or faulty internal grounding. Epoch Operator’s Manual...
Page 76
The gain/reset data for one or more wavelengths is missing. This data is collected during factory calibration and saved in nonvolatile memory. An error indicates that the internal software may be corrupted or the hardware has failed. BioTek Instruments, Inc.
Page 77
Failed configuration checksum test (1 = UI processor, 2 = MC processor). Probable causes: One of the processors on the PCB contains a defective • corrupt flash memory. The basecode software and/or assays may need to be redownloaded. • Main PCB failure. Epoch Operator’s Manual...
Page 78
This error is typically used only for software development purposes. If it occurs, however, try turning the instrument off, waiting 30 seconds, and then turning it on again. If the error persists, contact BioTek TAC. 1F0r Overlap error in PC command definition.
Page 79
Verify that the external power supply is correctly • connected. If error is intermittent, replace the external power • supply and/or the main PCB. Epoch Operator’s Manual...
INVALID_READSETS—number of readsets defined invalid INVALID_MODE—invalid mode selection received ASSAY_DEF_NOT_SET—validated assay not defined INVALID_ROW_COL—invalid number of rows and columns INVALID_EVENT_COUNT—total events defined invalid INVALID_EVENT_TYPE—invalid event type received PLATE_START_LATE—missed start of plate mode event INVALID_WELL_DATAPTS—number of samples out of range BioTek Instruments, Inc.
Page 81
INVALID_PLATE_INTRVAL—plate mode kinetic interval out of range READ_START_LATE—missed start of read INTERVAL_START_LATE—missed start of kinetic interval setup INVALID_DIRECTION—initial row or column direction is invalid DEBUG_MODE—unit is in debug mode; self-test may not have been run Epoch Operator’s Manual...
Fatal Errors Fatal errors indicate conditions that cannot be resolved by user actions. The instrument will no longer function. If a fatal error is displayed, contact BioTek’s Technical Assistance Center for further instructions (refer to the Product Support & Service section in Chapter 1). {data} indicates a parameter in the second-lowest digit of the error code.
Page 83
Instrument Dimensions This section shows the location of the microplate carrier in reference to the exterior surfaces of the Epoch and the mounting holes on the bottom. Use the illustrations to facilitate system setup with a robotic instrument, such as the...
Page 84
| Appendix C: Instrument Dimensions Special alignment hardware is included in the BioStack BioStack users: alignment kit for the Epoch. Refer to the BioStack Operator’s Manual for instructions for alignment of the BioStack with the reader. BioStack alignment pegs Figure 1: Front view of reader Figure 2: Side view of reader BioTek Instruments, Inc.
Page 85
Instrument Dimensions | BioStack alignment pegs Figure 3: Bottom view of reader, illustrating mounting holes for aligning caps (for operation with BioStack) Epoch Operator’s Manual...
Page 86
| Appendix C: Instrument Dimensions BioTek Instruments, Inc.
Index absorbance plate test installing, 9 defining parameters, 31 hazards, x description, 31 inspecting the Epoch, 6 example, 35 Installation Qualification results, 35 Overview, 26 running, 33 intended use, ix troubleshooting, 35 internal voltage, x absorbance test plate liquid testing, 37...
Page 88
RMA. See Return Materials TAC. See Technical Assistance Authorization number Center safety symbols, xv Technical Assistance Center, 4 selecting an appropriate location, turning on the Epoch, 9 unpacking instructions, 6 shipping hardware user safety, xiv attaching, 12 voltage, x removing, 7...
Need help?
Do you have a question about the Epoch and is the answer not in the manual?
Questions and answers